KinTek SF-MiniMixer
Specifications
Pneumatic Drive: Air pressure drives the pneumatic cylinder to force the mixing of solutions through the observation cell, with a stop syringe to limit the volume expended and to trigger data collection at stop.
Syringe chamber and cell assembly: The syringe chamber is transparent and can be thermostatted by circulating water (provided separately). In addition, a chamber above the syringes can be purged to eliminate oxygen above the samples. Samples contact only glass, Teflon, Kel-F and PEEK.
Syringe volume and range: 2 ml standard. Syringes of 0.25, 0.5, and 1.0 ml are also available for variable ratio mixing.
Temperature range: A range of 0-50°C is maintained by circulating water bath (provided separately).
Dead Time: The dead time is only 3.5 milliseconds. Reaction rates greater than 400s-1 have been measured.
Sample Volumes: The minimum reaction volume is 50 µl of each reactant per shot plus approximately 400 µl to flush and load a new sample.
Data Collection and Analysis: Data collection can be interfaced to KinTek SF software using a 200 kHz 12-bit A/D converter (available separately) providing fast and accurate data collection. Using the KinTek software, data can be fit to 1, 2 or 3 exponentials or 1, 2 exponentials plus a linear phase by nonlinear regression using the Marquat Method.
Observation Cells: The standard observation cell is a 2 × 2 × 5 mm quartz cuvette with a 5 or 2 mm path length (selectable by orienting the holder in the instrument). Other sample configurations are available on request for an additional charge.
Computer Interface: A TTL pulse is provided from the instrument to initiate data collection using software provided by the user or from KinTek.
Power Requirements: 110 volts, 15 amps.
| Dimensions: | (Width × Depth × Height) and Weight: |
|---|---|
| Stopped-Flow: | (48 × 15 × 6 cm) 8 kg |
| Power Supply: | (18 × 20 × 8 cm) 5 kg |
Warranty
The KinTek MiniMixer Stopped-Flow is backed by a 2-year warranty against defects in the pneumatic drive and power supply. Valves and syringes are covered by a 90-day warranty.